Diathermocoagulator for surgical purposes for cutting and coagulation of biological tissues with a high frequency strum (443 kHz). The pressure of the device should be adjusted for the treatment of the entire spectrum of electrosurgical interventions. You can compete in surgery, gynecology, dermatology, dentistry, urology, laparoscopy and endoscopy. Also, know your needs for medical cosmetology in SPA complexes and beauty salons.
The attachment works in two modes: monopolar and bipolar. The monopolar mode is subdivided into its line into the mode of cutting without coagulation and cutting with coagulation.
Monopolar mode vikoristovuє neutral electrode, which is to blame for buti fixing on the body of the patient near the place of the operation, and the electrode for surgical insertion, which without interruption of tissue growth. The Danish regime has increased respect for safety technology.
Bipolar mode vicorist electrode type tweezers for coagulation or cutting a small piece of tissue. The neutral electrode is not reversed in this mode.
Features of the updated model:
1. Stable fatigue is set independently at the hour of work.
2. Digital indication of the set tension for mono and bipolar modes.
3. The maximum suppression is 400 W
4. The maximum tension in the bipolar mode at a nominal tension with a support of 50 0 m -180 W ± 20%
5. Maximum output power in monopolar mode at nominal voltage 200 Ohm - 250 W ± 20%
6. Peak output intensity 300 W.
7. Modulation frequency 20 ±1 kHz
8. Depth of modulation
- in mode 1: 30%,
- in mode 2: 60%.
9. The maximum voltage at the output of the generator without voltage in monopolar mode is 1 200 V.
10. The maximum voltage at the generator output without voltage in the bipolar mode is 500 V.
11. Connection of all working tools (monopolar electrode trim, bipolar tweezers, passive electrode) is connected on the front panel of the device, which ensures maximum efficiency during the working hours.
12. Passive electrode of bagatorase whisker made from stainless steel, which can be seen in bagatorase autoclaving.
13. Visual light indication of external tension both for mono and for bipolar modes:
- Yellow color: 0% to 84%
- Red color: 84% to 100%
14. A permanent sound signal for robots in different modes:
- In bipolar mode - constant signal
- In monopolar mode - re-partial signal
Technical characteristics
1. Tension of lifeline -220V ± 20%
2. Mering frequency -50 Hz
3. The frequency of high-frequency coliving is 443 ± 5 kHz
4. The maximum suppression is 400 W
5. The maximum output pressure in the bipolar mode at a nominal voltage with a support of 50 Ohm -180 W ± 20%
6. Maximum output power in monopolar mode at nominal input 200 Ohm - 250 W ± 20%
7. Peak output intensity 300 W.
8. Modulation frequency 20 ± 1 kHz
9. Depth of modulation in mode 1 - 30%, in mode 2 - 60%.
10. The maximum voltage at the generator output without voltage in monopolar mode is 1200 V.
11. The maximum voltage at the generator output without voltage in the bipolar mode is 500 V.
12. Vaga - 5.2 kg
13. Dimensions dimensions, no more - 320 mm x 300mm x 100mm
Contents of delivery
1. Control unit for diathermocoagulator DKX-250 1 pc.
2. Trimach 4 mm, FD-100F4 - 1 pc.
3. Bipolar bagatorasian tweezers Q008;
4. Foot pedal - 1 pc.
5. Change parts (active electrodes):
- WFM 6;
- WFM 4;
- WFM 10;
- WFM 12;
- Electrode-lower Knife;
- BD70 (at the same time with trim).
6. Fusible insert - 1 pc.
7. 1 x life cable
8. Passive electrode - 1 piece
9. Passport - 1 pc.
Warranty period 12 months. from the day of commissioning